= Booking rules (last update: 11th Nov, 2014) =
Elionix 7000 is a 100 kV acceleration electron beam lithography tool.
*Between the hours of 9 am and 6 pm weekdays, every user is allotted a bank of 3 hours of bookable time.
== Tool Specification ==
*You may distribute these 3 hours across several slots - three 1 hour slots or two 1.5 hour slots, etc
* Acceleration voltage -25,50,75,100 kV
*You may book only 3 hours on the system at a time (See the after hours exception)
* Minimum beam diameter - 1.8 nm
*This rolling bank system also means that once you use, say 1 hour of your allotted 3 hours, you can make a new 1 hour booking sometime later.
* Beam current - 100 pA to 40nA
*After hours exception: 9am-6pm weekdays are the normal hours and reservation outside these hours (after hours) are not governed strictly. However, if I see unfair usage of the tool in this period, I will intervene.
* Field Size - Min 75 um X 75 um  Max 600 um
<del>Only one booking allowed at a time between 9am-6pm on weekdays, no regulations outside of these hours. Last updated: 15 Apr 2014.</del>
* Minimum line width - 8nm, 75 um field
* Overlay accuracy - 30nm


= Manuals and Guides =
Other lithography tools at the [[Main Page|NBI cleanroom]]:
== Local Qdev versions - cheat sheet & detailed guide ==
* [[Elionix F-125]], 125 kV EBL at QuanTech
<div class="toccolours mw-collapsible mw-collapsed">
* [[Raith eLine]], 30 kV EBL
* [https://wiki.nbi.ku.dk/qdevwiki/ElionixCheatSheet Cheat sheet]
* [[Heidelberg µPG 501]] LED writer
<div class="mw-collapsible-content">
* [[Süss mask aligner]]
* [[Media:ElionixQDevManual.pdf|QDev Elionix manual]] &mdash; written before the upgrade (linux->windows GUI), last updated 2013-03-05. Source repository [http://qdev-data.fys.ku.dk/git/repositories/view/guides/elionix-guide/ here], upload new pdf versions to [[File:ElionixQDevManual.pdf]].
* [[Media:Elionix_Manual_v2.pdf|Manual from Jimmy and Hugh]] &mdash; pertains to Harvard system, last updated 4/9/09.


=== Official ===
== Important notes on the fire alarm ==
[[File:ElionixFireAlarmPanel.jpg|thumb|left|Fig. 1. Fire alarm panel in the Elionix room.]]
The Elionix rooms are fitted with an overly sensitive fire alarm. Here is what happens as the room fills with smoke or dust:
# A very loud ''pre''-alarm starts to sound.
# If the level of smoke increases further, an inert gas is released and the level of oxygen in the room drops to 15% of the normal value. You should leave the room if it gets to this point.


* After upgrade:
Unfortunately, the detector is too sensitive; it has been known to go off because someone swept the floors of the room next door. Therefore, you need to know how to disable the alarm before the inert gas is released if the ''pre''-alarm goes off.
** [[Media:ElionixOverlayExposure.pdf|Overlay Exposures (Quick Reference)]] &mdash; by Taichi when he was in DK.
# Locate the control panel for the fire alarm next to the outer door. See Fig. 1.
** [[Media:ElionixGUITraining.pdf|Lose Ends from Training Session]] &mdash; by Taichi when he was in DK.
# Turn the left key to the <code>Manual Only</code> position to disable the release of the inert gas.
# Turn the middle key labeled <code>Enable Control</code> to unlock the control panel.
# Press the button labeled <code>Silence/Sound Alarm</code> to disable the horn.
# Press the button labeled <code>Alarm/Fault Warning Silence</code> to disable the buzzer.
# When the air is clean again, presse the button labeled <code>Reset</code> to reset the system. ''Then'' put all the keys back to their original positions.
# Double-check that the left key is in the <code>Auto & Manual</code> position. Otherwise the house will burn down if there is a fire.


* Before upgrade
== Booking rules ==
**[[Media:ElionixManualFull.pdf|Elionix 7000 User Manual (Full)]]
* Between the hours of 9 am and 6 pm weekdays, every user is allotted a bank of 3 hours of bookable time.
**[[Media:ElionixFieldCorrectionTips.pdf|Field correction tips]] &mdash; by Taichi when he was in DK.
* You may distribute these 3 hours across several slots - three 1 hour slots or two 1.5 hour slots, etc.
**[[Media:ElionixTrainingSummary.pdf|Training summary]] &mdash; by Taichi when he was in DK.
* Please be mindful of other users who might want to use the system for longer in-between your bookings.
**[[Media:ElionixApertureAdjust.pdf|Aperture initialization and adjustment]] &mdash; by Taichi when he was in DK.
* You may book only 3 hours on the system at a time (see the after hours exception).
**[[Media:ElionixStopStartManual.pdf|System shutdown and start-up manual]] &mdash; by Taichi when he was in DK.
* This rolling bank system also means that once you use, say 1 hour of your allotted 3 hours, you can make a new 1 hour booking sometime later.
* After hours exception: 9am-6pm weekdays are the normal hours and reservation outside these hours (after hours) are not governed strictly.
*: However, if you plan to use the tool overnight, please start as late as possible out of consideration for other users.
* Cleanroom staff reserves the right to intervene in case of booking system abuse.


</div>
== Operating instructions and troubleshooting ==
</div>
Currently hosted on the [https://wiki.nbi.ku.dk/qdevwiki/ElionixCheatSheet internal QDev wiki.]


= Design PC =
== How to transfer Exposure Files ==
This is the PC in the Elionix room in the basement, next to the two racks with elionix hardware. You may use this PC to ready your design, convert to *.cell & *.con formats. The desktop has all the shortcuts to network folders (Design computers on the 4th floor, shared images from SEM PC and user folder from Exposure PC) and relevant software. Any files outsides the designated user folders (shortcut on desktop) can be deleted without warning. The login credentials are:
: Username: ''Nanouser''
: Password:


The PC is labeled ''Elionix''.
1) Upload your files in N drive under the folder - " Elionix 100 Exp Files " ( Make sure to create your own folder )
 
2) Access your files in Elionix Exposure PC and copy them to local users folder .
Note:
* Only upload exposure files to the folder .
* Please do not access other general file from tool PC  
* You will find the shortcuts in Desktop
 
 
 
 
 
=== Remote access ===
Remote access to the tool is only granted to cleanroom staff.
 
 
 
[[Category:Tools]]
[[Category:Lithography]]

Latest revision as of 10:38, 9 March 2023

Elionix 7000
Picture of Elionix 7000 text
Essentials
Full nameElionix 7000
ManufacturerELIONIX INC.
DescriptionElectron beam lithography system
Location03.01.K03
Responsibility
PrimarySmitha
SecondaryZhe

Elionix 7000 is a 100 kV acceleration electron beam lithography tool.

Tool Specification

  • Acceleration voltage -25,50,75,100 kV
  • Minimum beam diameter - 1.8 nm
  • Beam current - 100 pA to 40nA
  • Field Size - Min 75 um X 75 um Max 600 um
  • Minimum line width - 8nm, 75 um field
  • Overlay accuracy - 30nm

Other lithography tools at the NBI cleanroom:

Important notes on the fire alarm

Fig. 1. Fire alarm panel in the Elionix room.

The Elionix rooms are fitted with an overly sensitive fire alarm. Here is what happens as the room fills with smoke or dust:

  1. A very loud pre-alarm starts to sound.
  2. If the level of smoke increases further, an inert gas is released and the level of oxygen in the room drops to 15% of the normal value. You should leave the room if it gets to this point.

Unfortunately, the detector is too sensitive; it has been known to go off because someone swept the floors of the room next door. Therefore, you need to know how to disable the alarm before the inert gas is released if the pre-alarm goes off.

  1. Locate the control panel for the fire alarm next to the outer door. See Fig. 1.
  2. Turn the left key to the Manual Only position to disable the release of the inert gas.
  3. Turn the middle key labeled Enable Control to unlock the control panel.
  4. Press the button labeled Silence/Sound Alarm to disable the horn.
  5. Press the button labeled Alarm/Fault Warning Silence to disable the buzzer.
  6. When the air is clean again, presse the button labeled Reset to reset the system. Then put all the keys back to their original positions.
  7. Double-check that the left key is in the Auto & Manual position. Otherwise the house will burn down if there is a fire.

Booking rules

  • Between the hours of 9 am and 6 pm weekdays, every user is allotted a bank of 3 hours of bookable time.
  • You may distribute these 3 hours across several slots - three 1 hour slots or two 1.5 hour slots, etc.
  • Please be mindful of other users who might want to use the system for longer in-between your bookings.
  • You may book only 3 hours on the system at a time (see the after hours exception).
  • This rolling bank system also means that once you use, say 1 hour of your allotted 3 hours, you can make a new 1 hour booking sometime later.
  • After hours exception: 9am-6pm weekdays are the normal hours and reservation outside these hours (after hours) are not governed strictly.
    However, if you plan to use the tool overnight, please start as late as possible out of consideration for other users.
  • Cleanroom staff reserves the right to intervene in case of booking system abuse.

Operating instructions and troubleshooting

Currently hosted on the internal QDev wiki.

How to transfer Exposure Files

1) Upload your files in N drive under the folder - " Elionix 100 Exp Files " ( Make sure to create your own folder )

2) Access your files in Elionix Exposure PC and copy them to local users folder .

Note:

  • Only upload exposure files to the folder .
  • Please do not access other general file from tool PC
  • You will find the shortcuts in Desktop



Remote access

Remote access to the tool is only granted to cleanroom staff.