Recommended parameters for EBL

From cleanroom
Jump to navigation Jump to search

Cleanroom recommendations

125 kV
Field Size (µm) Dots Pitch Pixel size (nm) Aperture (µm) Beam current (A) Beam spot (nm)
500 50000 8 80 240 60e-9 83.3
500 1000000 5 2.5 60 1e-9 2.6
100 kV
Field Size (µm) Dots Pitch Pixel size (nm) Aperture (µm) Beam current (A) Beam spot (nm)
600 20000 7 210 250 20e-9 218.6
600 240000 1 2.5 40 500e-12 2.7

How to choose the correct parameters yourself

A good place to start is the Unofficial dose calculator

  • Choose the EBL tool (100 or 125 keV)
  • Write field size (µm) and Dots: The combination of these two will give you the pixel size for a default pitch of 1.
    • It is recommended to use high number of dots with a pitch multiplier for fine exposures
  • Aperture and Beam current: The combination of these two will give you the beam spot size

The pixel size should be roughly equal to the beam spot size

  • Area dose: The recommended dose for the combination of substrate and resist stack that you have. If you don't know this value you can do a dose test for your substrate/resist combination or consult the Doses page