E-Gun evaporator
Overview
The e-gun evaporator, also known as Egon, is a homegrown electron beam physical vapor deposition system.
Operating instructions
Unloading sample/Venting loadlock
Prerequisites:
- Sample holder must be in the top position
- The gate valve (GV) must be closed
- The sample valve (SV) to the roughing pump must be closed
Then:
- Open vent valve on loadlock (LL). The pressure gauge for the LL should now read about 760 Torr or or (over range)
- Remove the sample holder
Loading sample/pumping down the loadlock
- Mount the sample on the holder and secure the holder in the top position
- Close the vent valve
- Close the roughing pump valve (RV). A warning signal for close RV starts
- Open the sample valve (SV). Watch the LL pressure on the gauge.
- When the LL pressure is < 8×10-2 torr, open RV again
- When the LL pressure is < 5×10-2 torr, close SV. Watch for a fast rise in the LL pressure! This is a sign of a leak.
- Open GV before LL pressure rises to > 8×10-2 torr
- Gently lower the sample holder into the chamber
Deposition
- Chamber pressure must be < 1×10-6 torr before starting
- Choose material on index wheel on the right side of the chamber
- Turn on the power supply (PS); Push all breakers down from left to right