Bruker Dimension Icon AFM: Difference between revisions
		
		
		
		Jump to navigation
		Jump to search
		
No edit summary  | 
				No edit summary  | 
				||
| Line 4: | Line 4: | ||
|company = Bruker  | |company = Bruker  | ||
|description = Atomic Force Microscope  | |description = Atomic Force Microscope  | ||
|location =   | |location = 03.K.08  | ||
|primary = Martin  | |primary = Martin  | ||
|secondary = Karolis  | |secondary = Karolis  | ||
| Line 21: | Line 21: | ||
*: - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)  | *: - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)  | ||
*: - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.  | *: - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.  | ||
[[Category:Tools]]  | |||
[[Category:Characterization]]  | |||
Revision as of 19:43, 6 April 2022
| Essentials | |
|---|---|
| Full name | Dimension Icon PT | 
| Manufacturer | Bruker | 
| Description | Atomic Force Microscope | 
| Location | 03.K.08 | 
| Responsibility | |
| Primary | Martin | 
| Secondary | Karolis | 
Quick start - ScanAsyst mode
- Open the enclosure hood. Red key goes counter-clockwise to unlock.
 - From the desktop, open the the Nanoscope software. This also powers up the tool (red lamp lit on scan head) .
 - Load the ScanAsyst in Air experiment.
 - In the left panel, select Setup.
- - In panel 1.1 make sure that the ScanAsyst-Air probe is selected.
 - - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window.
 - - Perform laser and photodetector alignment following steps 1.2-1.5.
 
 - In the left panel, select Navigate
- - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)
 - - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.