Bruker Dimension Icon AFM: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
|||
Line 15: | Line 15: | ||
* Load the ''ScanAsyst in Air'' experiment. | * Load the ''ScanAsyst in Air'' experiment. | ||
* In the left panel, select ''Setup''. | * In the left panel, select ''Setup''. | ||
*: In panel 1.1 make sure that the ''ScanAsyst-Air'' probe is selected. | *: - In panel 1.1 make sure that the ''ScanAsyst-Air'' probe is selected. | ||
*: Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window. | *: - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window. | ||
*: Perform laser alignment alignment following steps 1.2-1.5. | *: - Perform laser alignment alignment following steps 1.2-1.5. | ||
* In the left panel, select ''Navigate'' | * In the left panel, select ''Navigate'' |
Revision as of 12:12, 24 September 2020
![]() | |
Essentials | |
---|---|
Full name | Dimension Icon PT |
Manufacturer | Bruker |
Description | Atomic Force Microscope |
Location | K08 |
Responsibility | |
Primary | Martin |
Secondary | Karolis |
Quick start - ScanAsyst mode
- Open the enclosure hood. Red key goes counter-clockwise to unlock.
- From the desktop, open the the Nanoscope software. This also powers up the tool (red lamp lit on scan head) .
- Load the ScanAsyst in Air experiment.
- In the left panel, select Setup.
- - In panel 1.1 make sure that the ScanAsyst-Air probe is selected.
- - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window.
- - Perform laser alignment alignment following steps 1.2-1.5.
- In the left panel, select Navigate