Bruker Dimension Icon AFM: Difference between revisions

From cleanroom
Jump to navigation Jump to search
No edit summary
No edit summary
Line 11: Line 11:


== Quick start - ScanAsyst mode ==
== Quick start - ScanAsyst mode ==
* Open the enclosure hood. Key goes counter-clockwise to unlock.
* Open the enclosure hood. Red key goes counter-clockwise to unlock.
* From the desktop, open the the ''Nanoscope'' software. This also powers up the tool (red lamp lit on scan head) .
* From the desktop, open the the ''Nanoscope'' software. This also powers up the tool (red lamp lit on scan head) .
* Load the ''ScanAsyst in Air'' experiment.
* Load the ''ScanAsyst in Air'' experiment.

Revision as of 12:08, 24 September 2020

Bruker Dimension Icon AFM
Picture of Bruker Dimension Icon AFM text
Essentials
Full nameDimension Icon PT
ManufacturerBruker
DescriptionAtomic Force Microscope
LocationK08
Responsibility
PrimaryMartin
SecondaryKarolis

Quick start - ScanAsyst mode

  • Open the enclosure hood. Red key goes counter-clockwise to unlock.
  • From the desktop, open the the Nanoscope software. This also powers up the tool (red lamp lit on scan head) .
  • Load the ScanAsyst in Air experiment.
  • In the left panel, select Setup.
    In panel 1.1 make sure that the ScanAsyst-Air probe is selected.
    Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window.
    Perform laser alignment alignment following steps 1.2-1.5.
  • In the left panel, select Navigate