= Overview =
The e-gun evaporator, also known as [https://en.wikipedia.org/wiki/Egon_Olsen Egon], is a homegrown electron beam physical vapor deposition system.
= Operating instructions =
== Unloading sample/Venting loadlock ==
Prerequisites:
# Sample holder must be in the top position
# The gate valve (GV) '''must be closed'''
# The sample valve (SV) to the roughing pump '''must be closed'''
Then:
# Open vent valve on loadlock (LL). The pressure gauge for the LL should now read about 760 Torr or ''or'' (over range)
# Remove the sample holder

Revision as of 13:24, 2 December 2016

Template:Infobox

Overview

The e-gun evaporator, also known as Egon, is a homegrown electron beam physical vapor deposition system.

Operating instructions

Unloading sample/Venting loadlock

Prerequisites:

  1. Sample holder must be in the top position
  2. The gate valve (GV) must be closed
  3. The sample valve (SV) to the roughing pump must be closed

Then:

  1. Open vent valve on loadlock (LL). The pressure gauge for the LL should now read about 760 Torr or or (over range)