Bruker Dimension Icon AFM: Difference between revisions

From cleanroom
Jump to navigation Jump to search
No edit summary
No edit summary
Line 4: Line 4:
|company = Bruker
|company = Bruker
|description = Atomic Force Microscope
|description = Atomic Force Microscope
|location = K08
|location = 03.K.08
|primary = Martin
|primary = Martin
|secondary = Karolis
|secondary = Karolis
Line 21: Line 21:
*: - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)
*: - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)
*: - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.
*: - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.
[[Category:Tools]]
[[Category:Characterization]]

Revision as of 19:43, 6 April 2022

Bruker Dimension Icon AFM
Picture of Bruker Dimension Icon AFM text
Essentials
Full nameDimension Icon PT
ManufacturerBruker
DescriptionAtomic Force Microscope
Location03.K.08
Responsibility
PrimaryMartin
SecondaryKarolis

Quick start - ScanAsyst mode

  • Open the enclosure hood. Red key goes counter-clockwise to unlock.
  • From the desktop, open the the Nanoscope software. This also powers up the tool (red lamp lit on scan head) .
  • Load the ScanAsyst in Air experiment.
  • In the left panel, select Setup.
    - In panel 1.1 make sure that the ScanAsyst-Air probe is selected.
    - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window.
    - Perform laser and photodetector alignment following steps 1.2-1.5.
  • In the left panel, select Navigate
    - Load sample on the unblocked vacuum hole and toogle on the chuck vacuum by the switch (picture)
    - Drive the stage to place the sample under the scan head by using the XY navigation panel in the software or the track ball.