*: - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window. | *: - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window. | ||
*: - Perform laser | *: - Perform laser and photodetector alignment following steps 1.2-1.5. | ||
* In the left panel, select ''Navigate'' | * In the left panel, select ''Navigate'' | ||
Revision as of 12:13, 24 September 2020
| Essentials | |
|---|---|
| Full name | Dimension Icon PT |
| Manufacturer | Bruker |
| Description | Atomic Force Microscope |
| Location | K08 |
| Responsibility | |
| Primary | Martin |
| Secondary | Karolis |
Quick start - ScanAsyst mode
- Open the enclosure hood. Red key goes counter-clockwise to unlock.
- From the desktop, open the the Nanoscope software. This also powers up the tool (red lamp lit on scan head) .
- Load the ScanAsyst in Air experiment.
- In the left panel, select Setup.
- - In panel 1.1 make sure that the ScanAsyst-Air probe is selected.
- - Make sure that the cantilever is visible in the microscope. If you are not sure, zoom all the way out with the controls below the microscope window.
- - Perform laser and photodetector alignment following steps 1.2-1.5.
- In the left panel, select Navigate