AJA service
Jump to navigation
Jump to search
AJA systems maintenance is described here
Maintenance
Standard maintenance
- Log cryo temp, base pressure.
- Close cryo gate valve.
- Turn off ion gauge.
- Open loadlock gate valve.
- Turn off loadlock turbo.
- While venting:
- Unscrew right port with the linear crucible drive using two 9/16" wrenches.
- Unscrew lid if you intend to open it.
- Above 1e0 Torr loadlock pressure slowly open the vent nitrogen needle valve in front of the chamber.
- At atmosphere: start stopwatch.
- Pull out right port, open e-beam shutter.
- For each crucible:
- Wipe target metal surface with wipe
- Weigh with digital scale
- W crucible weighs ~120 g
- Intermetallic crucible ~20 g
- FabMate crucible ~12 g
- Gold pellets <80 g, 40-45 g for half
- Al pellets 6-7 g
- Top up target material if needed, log amount.
- If Sensor Life < 70% change the QCM. You can do this by sticking your hand through the loadlock. Be careful not to touch the mirror.
- AJA1: gold plated 6 MHz. There are two. Sensor 2 is towards the end of the assembly. It is a bit tricky to get out, even with the sensor shutter open.
- AJA2: silver plated 6 MHz (doesn't fail immediately during Pt evaporation)
- Push the linear drive back inside, screw the nuts back on the bolts
- Check if you can see the crucible in the mirror. If not:
- Attach chain to only one lid hook. Pull with ceiling motor for 20 mins
- Once the lid is open then lower the lid and lift with all three hooks
- Check if the quarter silicon wafer has started to delaminate. If so, peel off and reuse if possible. If it is not reflective enough, replace with clean quarter wafer.
- Adjust the clamp holding the mirror so that you can see the crucible through the port with LED light.
- Close lid, do not tighten screws/nuts
- Start pumping, log time at atmosphere, tighten the nuts on the crucible linear drive
- Close vent needle valve (not too tight!)
- If loadlock pressure goes below 1e-1 Torr, there are no obvious leaks
- After 30-60 mins check pressure. If <1e-4 Torr, turn on ion gauge
- Open cryo gate valve at own discretion. Turbo helps pumping down to ~1e-6 Torr. At lower pressures loadlock gate valve should be closed and only cryo should be open to the main chamber.
- Write a message to the users!
Cryo pump regeneration
- Close cryo VAT valve ( Adaptive pressure controller ) .
- Open the loadloack valve.
- Open Turbo is pumping the main chamber
- Turn off cryo (on the wall behind the tool). Just press and release switch .
- Open nitrogen vent valve on the back left of the cryo.
- Let warm to max T (~285 K), takes ~1 hour
- Close nitrogen vent valve
- Close roughing pump to turbo (screw valve under the loadlock turbo)
- Open roughing to cryo (screw valve next to cryo)
- Wait 20 mins to rough pump cryo
- Open roughing to turbo
- Turn on cryo ( wall switch) . Please check compressor as well .
- Wait ~60 mins to get between 200-150 K
- Close roughing to cryo
- Wait until min T. Must be below 20 K (2-3 hours at 2 K/min)
- If does not go down below 20 K, replace cryo with spare unit. Return cryo for repair/refurb to Edwards Vacuum. Contacts on the internal QDev wiki
- Close loadlock
- Open cryo VAT valve
Phase II J software configurations
The standard Phase II J software configurations on the systems are:
Troubleshooting
- Unable to ignite the plasma (either DC, RF or ion plasma)?
-
- Start by checking for shorts between pins on the power supply input, on the sputtering arm/ion source.
- No rate?
-
- E-beam shutter open?
- Correct material selected on deposition controller?
- Enough current?
- Beam in center of crucible and hitting the material?
- Enough material in the crucible?
- Rate falling during deposition?
- Material running out. Needs top-up.
- Crucible drive getting stuck?
-
- Drive support shafts worn + linear bearings gunked up. Replace
- Protect by wiping with IPA
- and covering with Al foil
- Crucible liner sideways. Abort and open system.
- Drive support shafts worn + linear bearings gunked up. Replace
- Can't turn on HV on the remote emission controller?
-
- Clear yellow error
- Make sure power supply is on
- Reconnect the controller cable
- Login as service: "2013"
- Adjust max emission so that 1% = 2.0/2.5 mA
- Can't adjust current?
-
- Left in auto mode. Change back to manual
- Emission knob encoder broken. Send back to factory to repair/replace.
- Red LED on loadlock gauge?
- Power cycle should fix it during next vent/pump.
- Lots of reflected power for an RF power source?
-
- The matching network for RF3 on AJA2 sometimes needs a bigger kick.
- Try turning on the power without ramping it.
- Another option would be to set the matching network to manual mode and strike the plasma, then turn it back to manual mode.
- The RF1 cable in the red shroud on AJA2 sometimes gets loose:
- Abort process, turn off RF milling power supply, reconnect cable, tighten as much as possible.
- The RF1 power supply controls are very sensitive, maybe someone touched them?
- Extremely slowly adjust Load to reach minimum of reflected power. If not 0 W, adust Tune. Iterate until 0 W.
- The matching network for RF3 on AJA2 sometimes needs a bigger kick.
- Recipes failing when adjusting gas flow?
- Adjust MFC timeout to 30 s
- user: service
- Software empty?
- Fill in with parameters the Phase II J software configurations.
- user: apex
Old notes
Older ion milling notes
When operating normally, the chamber should light a clear whiteish hue, and the kaufman power source should read numbers similar to these:

Service/repair/purchasing
More notes and service/repair/purchasing contacts on the internal QDev wiki.