How to load it
How to load it


===initialise settings and calibrate===
===Initialise settings and calibrate===
* Place the tape ring on the stage. ''Slide the ring underneath the two black tabs and ensure that it is sitting flush on the bottom bezel of the stage''
* Place the tape ring on the stage. ''Slide the ring underneath the two black tabs and ensure that it is sitting flush on the bottom bezel of the stage''
* Rotate the retaining lever to secure the tape ring.
* Rotate the retaining lever to secure the tape ring.
Line 59: Line 59:
** Is it aligned with the crosshair?
** Is it aligned with the crosshair?
*** '''Yes''' all good
*** '''Yes''' all good
*** '''No''' adjust the croshair by pressing <code>Recticle</code> in the correct x-direction so that it matches the peck line
*** '''No''' adjust the croshair by pressing <code>Recticle</code> in the correct x-direction (<code>-></code> or <code><-</code>)so that it matches the peck line
*** '''Can't see the peck line at all?''' Bring the tool a bit more down by rotating the wheel of the previous steps further. Then, repeat pecking.
*** '''Can't see the peck line at all?''' Bring the tool a bit more down by rotating the wheel of the previous steps further. Then, repeat pecking.
* If you want to save the sample position that you are in right now, press right click on <code>XV2</code> and <code>XV1</code> to store the positions.
* If you want to save the sample position that you are in right now, press right click on <code>XV2</code> and <code>XV1</code> to store the positions.
Line 66: Line 66:


===Scribe single lines===
===Scribe single lines===
theta adjustment
* Press <code>Two point up</code>
diec on x and Y aexe
diec on x and Y aexe
===Break===
===Break===


=== Remove chips, unload and cleanup===
=== Remove chips, unload and cleanup===

Revision as of 13:40, 7 July 2025

Loomis scriber
Picture of Loomis scriber text
Essentials
Full nameLoomis LSD-155
ManufacturerLoomis Inc
DescriptionAutomated semiconductor scriber and breaker
Location1st floor lab (03.1.111)
Responsibility
PrimaryHarry
SecondaryMartin

Other tools at the NBI cleanroom for wafer processing:

About

The Loomis LSD-155Lt is a production scribe and break system that can be used for processing large grids, arrays of laser bars, cleaving high quality mirror facets, and dicing wafers.

Primarily, it is dedicated to scribe and dice III/V samples, but it is also possible to use it for Si.

You can read the news about the tool in the designated Teams of UCPH_NBI_Cleanroom called Loomis

Policy

Only a limited number of users are trained on this tool. If you want to dice a wafer using this tool, contact cleanroom@nbi.ku.dk.]. Please include the following details:

  • Amount, material and crystal orientation of wafers.
  • Dicing dimensions
    • Is your wafer patterned? A schematic is well appreciated
  • Resist that the sample is coated

Standard operating procedure

Load your sample to the membrane

How to load it

Initialise settings and calibrate

  • Place the tape ring on the stage. Slide the ring underneath the two black tabs and ensure that it is sitting flush on the bottom bezel of the stage
  • Rotate the retaining lever to secure the tape ring.
  • Press Load Wafer
  • Press Load Recipe and select Si 001.
  • Choose the right tool

What is a tool?

A tool is practically a diamond tip with specific cuts that can scribe "deeper" or "broader" a surface.

  • Press Change tool
  • Find an area that you are fine if it gets damaged and press Tool down
  • If Tool touch indicates TOOL OFF then:
    • Rotate the wheel behind the microscope lens counterclockwise. **
    • While rotating the indication will change to TOOL ON. At that point, stop turning the wheel.
  • Press Tool up
  • Press Peck once
  • Press Tool up
  • Check the peck line:
    • Is it aligned with the crosshair?
      • Yes all good
      • No adjust the croshair by pressing Recticle in the correct x-direction (-> or <-)so that it matches the peck line
      • Can't see the peck line at all? Bring the tool a bit more down by rotating the wheel of the previous steps further. Then, repeat pecking.
  • If you want to save the sample position that you are in right now, press right click on XV2 and XV1 to store the positions.

Right click is storing, left click is going

Scribe single lines

  • Press Two point up

diec on x and Y aexe

Break

Remove chips, unload and cleanup