== Weight calibration ==
== Weight calibration ==
''adapted from an original SOP by Masoomeh''
=== Procedure ===
=== Procedure ===
* Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g.
* Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g.

Revision as of 12:56, 17 January 2022

FS bonder
Picture of FS bonder text
Essentials
Full nameF&S Bondtec 5630
ManufacturerF&S Bondtec
DescriptionAutomatic bonder
Location1st floor lab (03.1.111)
Responsibility
PrimaryNader
SecondaryKarolis

Standard operating procedure

adapted from an original SOP by Masoomeh

This is a standard operating procedure (SOP) describing the use of F&S 5630 bonder on the first floor Lab.

Preparation

  • Turn on the light on top of the laminar air flow bench. Set flow between 2-5.
  • Open the sash.
  • Start the 5630 program [red]
    • Do you want to move home?: Yes
    • Do you want to open the last bonding program?: No
  • Load the appropriate block for your daughterboard. There are three different blocks corresponding to different daughterboards/PCBs:
    • Copenhagen
    • Sydney
    • transmon
  • Remove the dummy daughterboard from the block.
  • Screw your PCB (with a chip glued onto it) onto the block.
    • In case of the transmon block, you can further secure your board to the block:
      • Turn on the vacuum pump located in the left behind the bonder
      • Click on "device clamp" in ribbon.

First-time setup

  • File > New
    • "Create Program Dialog" box appears.
      • Data source: Create from default value
        • 25 um
  • Use the joystick to move to the bonding pads on the daughterboard (not your chip)
    • It should be at least 5 mm away from any screws or connectors.
  • "Global Camera setting": Adjust Gain and Shutter to make the daughter board visible.
  • Adjust head height to see the PCB in focus
    • press button on joystick, tilt joystick down slowly
  • Next
  • Store trace - 6 to set "Workheight1"
  • Store trace - 6 to set "Workheight2"
  • Bring the bonding head halfway down (no closer than 5 mm above the highest point of any screws)
  • Store trace - 6 to set "Joystick Limit"
  • Store trace - 6 to set "max z position"
    • The head will touch down on the PCB and adjust the height. Note that the head might leave a small mark where it touches down.
    • In all of these four states you don’t need to move the stage or change the height of the head (except height for "joystick limit").
  • Next
  • Bonding parameter suggestions:
    • Bond force: 25
    • US power: 80
    • Total bond time: 25
    • Other parameters default
      • You can change these parameters later ant any point before/during/after bonding.
      • You are encouraged to look up parameters that previous users have used on similar surfaces.
        • The easiest way to gain access to other software at this point is to press Windows key.
  • File > Save as... to save your bonding program.
    • Please make a folder for yourself in the users folder.

Bonding

  • File to load your program.
  • F1 to bring up the bonding dialog.
  • In Bond1:
    • Move the stage near the bonding pads of your PCB, at least 5 mm away from connectors or screws.
    • Adjust light settings and focus on the PCB.
    • Store trace - 6 to store PCB height.
  • Press Right - 2 or click on Bond2. Note that the head might leave a small mark where it touches down.
  • In Bond2:
    • Move the stage near the bonding pads of your chip.
    • Adjust light settings and focus on the chip.
    • Store trace - 6 to store chip height. Note that the head might leave a small mark where it touches down.
  • Now you are ready to bond! Drag and drop Bond1 and Bond2 on the desired positions and then press F1 to bond.

Notes:

  • Arrows left/right move between Bond1 and Bond2
  • Pressing F1 bond on Bond1 moves to Bond2 position
  • Pressing F1 bond on Bond2 bonds
  • Pressing 0 home | escape homes the stage, turns off light, turns off camera

Clean-up

  • Check the thread
    • Feed # feeds a bit of the wire. If the wire does not come out of the wedge you need to rethread the wire.
  • Close the bonding window. Save the bond program by selecting [File > Save] if you would like to save your changes.
  • Close the 5630 program.
  • Wait until the sample stage move to its home position.
  • Unscrew your PCB.
  • Check that all of the fuzz buttons are present. If not, replace.
  • Screw on the protector plate.
  • Fill in the log. Requires fields are name, date and comments/problems if there are any.
  • Turn off the vacuum pump if you were using it during bonding.
  • Close the sash.
  • Turn down the air flow to 1.
  • Turn off the light.


Weight calibration

adapted from an original SOP by Masoomeh

Procedure

  • Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g.
  • Make a new bond program for the height of the scale.
  • Bonder software: Mode > Head calibration
  • In BW-Calib tab:
    • Weight point 1: Store trace - 6
      • Needle comes down and presses on the scale.
      • Update weight point 1.
    • Weight point 2: Store trace - 6
      • Needle comes down and presses on the scale.
      • Update weight point 2.
    • Repeat until both weights are accurate within 1 g.

Calibration test

  • Open the bond program for the scale.
  • F1 to open the bonding window.
  • Click on Bond1.
  • Set the bonding force to e.g. 25 g and US power to 0.
  • Change to step mode with F2.
  • F1 to go through the bonding steps until the head presses on the scale.
  • Verify that the weight is within 1 g, otherwise repeat the procedure.