== Weight calibration == | == Weight calibration == | ||
''adapted from an original SOP by Masoomeh'' | |||
=== Procedure === | === Procedure === | ||
* Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g. | * Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g. |
Revision as of 12:56, 17 January 2022
![]() | |
Essentials | |
---|---|
Full name | F&S Bondtec 5630 |
Manufacturer | F&S Bondtec |
Description | Automatic bonder |
Location | 1st floor lab (03.1.111) |
Responsibility | |
Primary | Nader |
Secondary | Karolis |
Standard operating procedure
adapted from an original SOP by Masoomeh
This is a standard operating procedure (SOP) describing the use of F&S 5630 bonder on the first floor Lab.
Preparation
- Turn on the light on top of the laminar air flow bench. Set flow between 2-5.
- Open the sash.
- Start the 5630 program [red]
- Do you want to move home?: Yes
- Do you want to open the last bonding program?: No
- Load the appropriate block for your daughterboard. There are three different blocks corresponding to different daughterboards/PCBs:
- Copenhagen
- Sydney
- transmon
- Remove the dummy daughterboard from the block.
- Screw your PCB (with a chip glued onto it) onto the block.
- In case of the transmon block, you can further secure your board to the block:
- Turn on the vacuum pump located in the left behind the bonder
- Click on "device clamp" in ribbon.
- In case of the transmon block, you can further secure your board to the block:
First-time setup
File > New
- "Create Program Dialog" box appears.
- Data source: Create from default value
- 25 um
- Data source: Create from default value
- "Create Program Dialog" box appears.
- Use the joystick to move to the bonding pads on the daughterboard (not your chip)
- It should be at least 5 mm away from any screws or connectors.
- "Global Camera setting": Adjust Gain and Shutter to make the daughter board visible.
- Adjust head height to see the PCB in focus
- press button on joystick, tilt joystick down slowly
Next
Store trace - 6
to set "Workheight1"Store trace - 6
to set "Workheight2"- Bring the bonding head halfway down (no closer than 5 mm above the highest point of any screws)
Store trace - 6
to set "Joystick Limit"Store trace - 6
to set "max z position"- The head will touch down on the PCB and adjust the height. Note that the head might leave a small mark where it touches down.
- In all of these four states you don’t need to move the stage or change the height of the head (except height for "joystick limit").
Next
- Bonding parameter suggestions:
- Bond force: 25
- US power: 80
- Total bond time: 25
- Other parameters default
- You can change these parameters later ant any point before/during/after bonding.
- You are encouraged to look up parameters that previous users have used on similar surfaces.
- The easiest way to gain access to other software at this point is to press
Windows key
.
- The easiest way to gain access to other software at this point is to press
File > Save as...
to save your bonding program.- Please make a folder for yourself in the users folder.
Bonding
File
to load your program.F1
to bring up the bonding dialog.
- In
Bond1
:- Move the stage near the bonding pads of your PCB, at least 5 mm away from connectors or screws.
- Adjust light settings and focus on the PCB.
Store trace - 6
to store PCB height.
- Press
Right - 2
or click onBond2
. Note that the head might leave a small mark where it touches down.
- In
Bond2
:- Move the stage near the bonding pads of your chip.
- Adjust light settings and focus on the chip.
Store trace - 6
to store chip height. Note that the head might leave a small mark where it touches down.
- Now you are ready to bond! Drag and drop Bond1 and Bond2 on the desired positions and then press
F1
to bond.
Notes:
- Arrows left/right move between Bond1 and Bond2
- Pressing
F1
bond on Bond1 moves to Bond2 position - Pressing
F1
bond on Bond2 bonds - Pressing
0 home | escape
homes the stage, turns off light, turns off camera
Clean-up
- Check the thread
Feed #
feeds a bit of the wire. If the wire does not come out of the wedge you need to rethread the wire.
- Close the bonding window. Save the bond program by selecting [File > Save] if you would like to save your changes.
- Close the 5630 program.
- Wait until the sample stage move to its home position.
- Unscrew your PCB.
- Check that all of the fuzz buttons are present. If not, replace.
- Screw on the protector plate.
- Fill in the log. Requires fields are name, date and comments/problems if there are any.
- Turn off the vacuum pump if you were using it during bonding.
- Close the sash.
- Turn down the air flow to 1.
- Turn off the light.
Weight calibration
adapted from an original SOP by Masoomeh
Procedure
- Place a digital weighing scale on the bonder sample holder. You will need a scale with a range over 200 g.
- Make a new bond program for the height of the scale.
- Bonder software:
Mode > Head calibration
- In BW-Calib tab:
- Weight point 1:
Store trace - 6
- Needle comes down and presses on the scale.
- Update weight point 1.
- Weight point 2:
Store trace - 6
- Needle comes down and presses on the scale.
- Update weight point 2.
- Repeat until both weights are accurate within 1 g.
- Weight point 1:
Calibration test
- Open the bond program for the scale.
F1
to open the bonding window.- Click on
Bond1
. - Set the bonding force to e.g. 25 g and US power to 0.
- Change to step mode with
F2
. F1
to go through the bonding steps until the head presses on the scale.- Verify that the weight is within 1 g, otherwise repeat the procedure.