Working with KOH: Difference between revisions
Jump to navigation
Jump to search
No edit summary |
No edit summary |
||
| Line 1: | Line 1: | ||
KOH belongs to the family of anisotropic Si-etchants based on aqueous alkaline solutions. The anisotropy stems from the different etch rates in different crystal directions. The {111}-planes are almost inert whereas the etch rates of e.g. {100}- and {110}-planes are several orders of magnitude faster. | KOH belongs to the family of anisotropic Si-etchants based on aqueous alkaline solutions. The anisotropy stems from the different etch rates in different crystal directions. The {111}-planes are almost inert whereas the etch rates of e.g. {100}- and {110}-planes are several orders of magnitude faster. | ||
At NBI cleanroom, we use as standard a 28 | At NBI cleanroom, we use as standard a 28% w/w KOH solution. The etch rate and the selectivity towards a SiO2 mask depends on the temperature. We normally use a temperature of 80°C but it may reduce it to, e.g., 60 °C or 70°C for a higher-precision timed etch. | ||
{| class="wikitable" | {| class="wikitable" | ||
| Line 23: | Line 23: | ||
| 80°C || 0.0075 | | 80°C || 0.0075 | ||
|- | |- | ||
|} | |||
=Standard Operating Procedure= | |||
==Mixing KOH (with pellets)== | |||
* Book the HF fumehood in CR1. | |||
: ''It is recommended to use a setup where you can run your process undisturbed'' | |||
* Gather all necessary materials and equipment: | |||
** IKA hotplate with 2L beaker | |||
** Scale from to shelves | |||
** Beakers for measuring and mixing | |||
** KOH pellets | |||
* Set up equipment | |||
** Fill 2L beaker with water to create a hotbath and add a magnetic stirring pill | |||
** Balance the scale | |||
* Calculate the portions of water and solid KOH pellets that you would like to mix: | |||
** You are aiming to make a 28% w/w solution | |||
: There is no shame on using an [https://www.physiologyweb.com/calculators/percent_solutions_calculator.html online calculator] to figure out the ratios. | |||
** The KOH pellets come in units of xxx grams. Start by weighing an appropriate amount of KOH and store it to the side. | |||
** Calculate the respective amount of water and fill a beaker with it. | |||
* Slowly add the KOH pellets to the water a few at a time while stirring gently. | |||
* Continue stirring until the pellets dissolve completely | |||
* Place your solution in the hotbath | |||
* Place a second beaker with MQ next to it | |||
* Submerge the probe heater into the MQ beaker | |||
* Start the stirring and set the temperature. | |||
* Once you reach the desired temperature, you can start etching your sample | |||
Revision as of 15:49, 9 September 2026
KOH belongs to the family of anisotropic Si-etchants based on aqueous alkaline solutions. The anisotropy stems from the different etch rates in different crystal directions. The {111}-planes are almost inert whereas the etch rates of e.g. {100}- and {110}-planes are several orders of magnitude faster.
At NBI cleanroom, we use as standard a 28% w/w KOH solution. The etch rate and the selectivity towards a SiO2 mask depends on the temperature. We normally use a temperature of 80°C but it may reduce it to, e.g., 60 °C or 70°C for a higher-precision timed etch.
| Si(001) temperature | Etch rate (μm/min) |
|---|---|
| 60°C | 0.4 |
| 70°C | 0.7 |
| 80°C | 1.3 |
| Si(110) temperature | Etch rate (μm/min) |
| 80°C | 2.5 |
| Thermal SiOx temperature | Etch rate (μm/min) |
| 60°C | 0.0012 |
| 80°C | 0.0075 |
Standard Operating Procedure
Mixing KOH (with pellets)
- Book the HF fumehood in CR1.
- It is recommended to use a setup where you can run your process undisturbed
- Gather all necessary materials and equipment:
- IKA hotplate with 2L beaker
- Scale from to shelves
- Beakers for measuring and mixing
- KOH pellets
- Set up equipment
- Fill 2L beaker with water to create a hotbath and add a magnetic stirring pill
- Balance the scale
- Calculate the portions of water and solid KOH pellets that you would like to mix:
- You are aiming to make a 28% w/w solution
- There is no shame on using an online calculator to figure out the ratios.
- The KOH pellets come in units of xxx grams. Start by weighing an appropriate amount of KOH and store it to the side.
- Calculate the respective amount of water and fill a beaker with it.
- Slowly add the KOH pellets to the water a few at a time while stirring gently.
- Continue stirring until the pellets dissolve completely
- Place your solution in the hotbath
- Place a second beaker with MQ next to it
- Submerge the probe heater into the MQ beaker
- Start the stirring and set the temperature.
- Once you reach the desired temperature, you can start etching your sample